二氧化氯ClO2氣體檢測(cè)管116 CHLORINE DIOXIDE 1-20ppm,常溫常壓混合條件下,瓶裝純凈水用臭氧消毒時(shí),通常1m3/h水需使用3g臭氧,并且水中臭氧濃度需達(dá)到或超過(guò)0.3mg/L;對(duì)瓶裝礦泉水臭氧消毒時(shí),通常1m3/h水需使用6g臭氧,并且水中臭氧濃度需達(dá)到或超過(guò)0.5mg/L。
更新時(shí)間:2025-01-08
在線留言二氧化氯ClO2氣體檢測(cè)管116 CHLORINE DIOXIDE 1-20ppm
Tube No.
116 CHLORINE DIOXIDE
1. PERFORMANCE
1) Measuring range :1-20 ppm
Number of pump strokes 1(100ml)
2) Sampling time :2 minutes/1 pump stroke
3) Detectable limit :0.3 ppm
4) Shelf life :2 years
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Concentration chart method
8) Colour change :White→Reddish orange
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:5% RSD-high:5%
3. CHEMICAL REACTION
0-Toluidine is oxidized and Orthoquinone is produced.
4. CALIBRATION OF THE TUBE
COLOURIMETRY METHOD
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump ??????????????????????????????????????
??????????????????????????????????????
Substance Interference ppm Coexistence
Nitrogen dioxide FIG.1 Similar stain is produced. 1 Higher readings are given.
Chlorine FIG.2 〃1 〃
Bromine FIG.3 〃1 〃